Quality Control Testing

Since concrete is a manufactured product, it requires specific control tests and evaluations during the manufacturing process and at the job site to ensure high-quality concrete. Some of the important concrete properties measured by basic quality control tests are strength, temperature, slump, air content, and unit weight. Each test helps to determine the quality of concrete and should be performed in accordance with American Society for Testing and Materials (ASTM) standards.

Sampling Fresh Concrete (ASTM C 172)
Concrete used for testing should be representative of the entire batch. A good practice is to take two or more portions from the middle-third of the truckload and combine the portions. Take samples after all water has been added. Obtain a minimum 2-cubic-foot sample size if cylinders will be made.

Slump Testing (ASTM C 143) measures the consistency of concrete and often reflects a change in the amount of mix water, temperature, setting, and hydration. This test should be done in the field.

Air Content (ASTM C 231) Testing measures the amount of air bubbles in fresh concrete. Entrained air is purposely added to concrete for freeze-thaw resistance. Too much air results in low strength. This test should be done in the field.

Strength Testing - Making Cylinders (ASTM C 31)
Cylinders are made and tested to determine compliance with strength specifications. Test specimens must be cast, handled, cured and tested in accordance with ASTM specifications dealing with both field and laboratory test specimens. Cylinders are made in the field and tested in the lab.

To make a cylinder:
1.
Place cylinder mold on a level, rigid, horizontal surface, free from vibration and other disturbances and as close as possible to the location where it will be stored for the first 24 hours.

2. Fill the mold in three equal layers.

3.  Rod each layer 25 times, penetrating into the previous layer about 1 inch. Tap side of the mold lightly after each lift to close voids.

4. Strike-off the surface and cover immediately to prevent evaporation.

5. Store cylinders made to check strength on site for 16 to 24 hours at a temperature of 60 to 80° F then ship to the testing Lab.

6. Protect cylinders from damage during storage and shipping.

Low Test Results

Low test results can occur for a number of reasons. Not following prescribed standard ASTM procedures can cause low results:
•  Cylinder and beam test results will represent only about 20% of the actual strength of concrete in place.
•  Certain non-standard conditions can reduce test strengths by as much as 75%. Non-Standard conditions comprise improper casting, handling, curing and testing both in the field and in the testing laboratory. (For a complete list, see "American Society of Civil Engineers Document, Vol. 3 No. 2 Table # 1, May 1991.")

Were the cylinder test results evaluated properly?
•  The average of two cylinder strengths made from the same sample of concrete are required for one individual test.
•  The average of any 3 consecutive tests is required to be above the specified strength.
•  No individual strength test should fall below the specified strength by more than 500 psi.
•  Strength test failing to meet the criteria will occur occasionally (probably about once in 100 tests). Allowances should be made for such statistically expected variations in deciding whether the strength level being produced is adequate.

Was it the approved mix that was delivered and sampled? You must determine:
•  Whether the right mix was ordered and dispatched.
•  Whether the mix was batched correctlycheck for possible omission of water reducer, excessive air entrainment, or low cement content.
•  Whether any unauthorized or unrecorded water was added on the jobsite.

 

Tips for Job Site Testing
• Insist that trained personnel perform all tests.

• Know the importance of each test.

• Require that all tests conform to ASTM standards.

• Look for trends in test results to head off problems.

• Don't confuse strength tests for meeting the strength specification with tests that only indicate strength of concrete in place.

• Don't let samples be taken that are not representative of the batch or load being tested.

• Don't let test specimens stand unprotected on the job.

• Don't wait for a test failure before pointing out nonstandard test procedures.

   
 


  © 2009 Prairie Material
HOME | READY MIX | DECORATIVE CONCRETE | PERVIOUS CONCRETE | AGGREGATES
LOGIN | SITE MAP | PRIVACY | TERMS & CONDITIONS